Automatic Viscometer Models CAV-2100 and CAV-2200 with VISCPRO
®
Instruction & Operation Manual
Version 2g — May, 2009; CANNON
®
Instrument Company
2139 High Tech Road • State College, PA 16803 • USA
The
View Instrument Group
window
7. Click OK to display the Sample Input View for the desired group:
The Sample Input window
Then click the Tray tab corresponding with the desired viscometer tube.
8. Double-click on Sample ID (1) with the left mouse button to access
the sample ID data entry field (or press 2).
9. Enter sample ID information in the sample list box using your
computer keyboard. After you have typed the sample ID, press the
T key to save the entry and move the cursor to the next Sample
ID field. Or press R to save the entry.