Mini-Rotary Viscometer CMRV-4500 Instruction & Operation Manual
Version 1.1e—October, 2011 • CANNON
®
Instrument Company
2139 High Tech Road • State College, PA • 16803 • USA
NOTE
Some fields/options are security-protected, and may not be viewable.
The View Cell Constants window
3. Click Configuration/Print Instrument Settings to open the Win-
dows
®
Print window. Select the desired printer and click OK to print
current calibration settings.
NOTE
CANNON recommends printing calibration settings each time the
calibration values change.
Setting multiple CMRV instrument addresses
When installing/connecting multiple CMRV instruments, ensure that the
instrument address for each instrument is different. See APPENDIX D
for more information on multi-unit configuration.
Security options
The VISCPRO
®
application offers a level-based security system with a
log in procedure. This system ensures that available software functions
are appropriate for the user's needs.
Following initial installation of the software, security options may be
accessed and changed from the VISCPRO
®
Main menu by individuals
who have the manager password (initially blank).
Each user can be assigned a security level, which is used to determine
permitted operations. Three security levels are defined by the software:
Technician (least privileged), Manager (intermediate privilege), and
Service (maximum privilege–reserved for CANNON
®
Instrument
Company service personnel).
Users log in when using the software by typing their name and (optional)
password in the Log In window. Different security levels offer different
instrument use options. Technicians are permitted to access operational