Instruction & Operation Manual
Version 1.a— January, 2011; CANNON
®
Instrument Company
2139 High Tech Road • State College, PA 16803 • USA
Aborting a test
To permanently halt testing for a given tube/tray, click the Abort button
from the Sample Input window. Then select the desired trays from the
Select Trays window. Click OK to abort testing for the selected tray.
NOTE
Aborting a test clears all sample test information for that tray. If test
actions are aborted, it is the responsibility of the user to restore the
instrument to a safe state before running tests (see Service menu options
in chapter 4 for more information on tube washing and drying).
Concluding a test
After automatic testing has been completed, make certain that:
1. the sample vial carrier is in the lowered position.
CAUTION
Use appropriate procedures when handling warm sample vials to avoid the
possibility of burns.
2. the Machine Status (as indicated in the VISCPRO
®
Sample Input
Machine Status window), is READY, and
3. the Tray Status for the tube, as indicated in the VISCPRO
®
Sample
Input Tray Status window, is IDLE. Kinematic viscosity for any
tested samples will be displayed in the Sample Action column in the
action list for the selected tube.
NOTE
To review status of other online instruments, click the tube tab corre-
sponding to the desired instrument (see next section).
If necessary, wipe any excess oil from the base of the miniPV Series,
sample vial and vial holder using an absorbent paper towel. If necessary,
clean these components before reuse by wiping with a paper towel
wetted with appropriate solvent.
Working with Instrument Groups
VISCPRO II provides a convenient interface for working with multiple
CANNON instruments simultaneously. Rather than requiring the user to
open individual Instrument Views for each online instrument, the Sample
Input window (formerly the Instrument View window in VISCPRO)
provides individual tabs for each viscometer tube, making it easy to enter
sample information for up to four different instruments. The Machine
Status window makes it possible to monitor the performance of all
instruments in an Instrument Group simultaneously.